SETTINGS
K0553 Revision D 52 [EN] English
- Pressure units
Shows the current selection. To change the pressure units setting:
1) Tap the pressure units panel.
2) Tap the required units radio button. The units radio button panel closes and the pressure
units panel shows the selected units.
- Aero units
Shows the current selection. To change the aero units setting:
1) Tap the aero units panel.
2) Tap the required aero units radio button. The aero units radio button panel closes and the
aero units panel shows the selected units.
- Altitude correction
Shows the current selection. To change the altitude correction setting:
1) Tap the altitude correction panel, a numbered keypad is displayed.NOTE: Enter a positive
correction factor when ADTS is below the aircraft reference. Enter a negative correction
factor when ADTS is above the aircraft reference.
2) On the numbered keypad, select the required altitude correction setting.
3) Tap the “Tick” icon, the numbered keypad closes and the new altitude correction setting is
displayed in the altitude correction setting panel.
4) Taping the “Cross” icon, cancels the action and closes the numbered keypad.
PS1-PS2 differential limit in 3 channel operation
The default value for the Ps1-Ps2 limit is 84.66 mbar (2.5 inHg) and can be changed via the hand
terminal.

Table of Contents

Other manuals for GE ADTS542F

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE ADTS542F and is the answer not in the manual?

GE ADTS542F Specifications

General IconGeneral
BrandGE
ModelADTS542F
CategoryTest Equipment
LanguageEnglish

Related product manuals