Chapter 2. Instrument Setup
USM Go Operator’s Manual 59
2.7 Using the Calibration Reminder Alarm
The USM Go incorporates a timed alarm feature that causes an icon to appear on user-defined input intervals between 0.5 and 4.0 hours. To
use this alarm:
1. In the SETUP menu, activate the CONFIG2 submenu using the joystick ( ).
2. Use the joystick ( ) to select the function titled CAL REMINDER, then press the center of the joystick ( ) to activate the
function. Input the intervals at which the reminder alarm will trigger (between 0.5 and 4.0 hours). Setting this value automatically resets
the alarm to trigger on the specified intervals. This function also allows the alarm to be disabled.
3. Press the function toggle or move the joystick ( ) to change the alarm interval.
4. Select and activate CAL RESET to acknowledge the triggered alarm and reset it to resume normal operation.
Note: The selection of the PROBE DELAY or VELOCITY functions also resets the alarm.