Chapter 3 - Setup Instructions 3-83
3-12-5 How to enter Patient's Email address and Phone number in the PID screen
1.) Enter the “Patient ID” screen by pressing the Patient ID key on the control console.
2.) Click the @ icon (1).
The Patient’s Email address and Phone number window appear.
3.) Enter patient's information and adjust Email/MMS Options (2).
4.) Store your configuration with click on OK (3).
NOTE: Since the E-mail to MMS service provider is maybe not located in the same area as your MMS
receptions, you need to add the area code (e.g. +49 or 0049 for Germany) to the used phone numbers
in the Patient's E-mail address and Phone number dialog.
Example:
local phone number 069913800000 is +49 69913800000 or 0049 69913800000 for Germany
Figure 3-74 Patient ID screen