Page #390 background imageLoading...
Page #390 background image
GE RAFT VOLUSON™ P8/VOLUSON™P6
DIRECTION 5775469, R
EVISION 3 DRAFT (JULY 19, 2018) BASIC SERVICE MANUAL
8-28 Section 8-11 - Replacement of Cover Top Cabinet Plastic
Section 8-11
Replacement of Cover Top Cabinet Plastic
8-11-1 Manpower
1 people, 3 minutes
8-11-2 Tools
Not required
8-11-3 Disassembly Procedure
1.) Remove the FRONT CABINET BEZEL ASSY BT16. Refer to the figure below.
2.) Remove the FRONT PRINTER BEZEL ASSY BT16. Refer to the figure below.
Figure 8-25 Removing the FRONT CABINET BEZEL ASSY BT16
Figure 8-26 Removing the FRONT PRINTER BEZEL ASSY BT16
FRONT CABINET BEZEL ASSY BT16
FRONT PRINTER BEZEL
ASSY BT16