GE DRAFT VOLUSON™ P8/VOLUSON™P6
DIRECTION 5775469, R
EVISION 3 DRAFT (JULY 19, 2018) BASIC SERVICE MANUAL
Chapter 8 - Replacement Procedures 8-111
Section 8-34
Replacement of the TGC BOARD
8-34-1 Manpower
One person, 10 minutes
8-34-2 Tools
Philips Screwdriver
8-34-3 Preparations
1.) Power Off/Shutdown the system as described in 4-3-2 on page 4-4.
8-34-4 Removal Procedure
1.) Follow the removal procedure of Section 8-25 "Replacement of OPIO" .
2.) Remove 8 TGC control key caps. refer to Section 8-19 "Replacement of OPIO Plastic lower cover
assy" on page 8-56
3.) Disconnect TGC BOARD CABLE. Refer to the figure below.
Figure 8-164 Disconnecting TGC BOARD CABLE