GE RAFT VOLUSON™ P8/VOLUSON™P6
DIRECTION 5775469, R
EVISION 3 DRAFT (JULY 19, 2018) BASIC SERVICE MANUAL
8-98 Section 8-29 - Replacement of the AIR FILTER
Section 8-29
Replacement of the AIR FILTER
8-29-1 Manpower
One person, 1 minutes
8-29-2 Tools
None
8-29-3 Preparations
1.) Power Off/Shutdown the system as described in 4-3-2 on page 4-4.
8-29-4 Removal Procedure
1.) Pull out the FRONT CABINET BEZEL ASSY BT16. Refer to the figure below.
2.) Pull out the AIR FILTER. Refer to the figure below.
Figure 8-145 Removing the FRONT CABINET BEZEL ASSY BT16
Figure 8-146 Removing the AIR FILTER