GE DRAFT VOLUSON™ P8/VOLUSON™P6
DIRECTION 5775469, R
EVISION 3 DRAFT (JULY 19, 2018) BASIC SERVICE MANUAL
Chapter 8 - Replacement Procedures 8-169
Section 8-55
Replacement of UVC
8-55-1 Manpower
One person, 30 minutes
8-55-2 Tools
Philips screwdriver
8-55-3 Preparation
1.) Power Off/Shutdown the system as described in 4-3-2 "Power Off / Shutdown" on page 4-4
2.) Remove the rear covers and disconnect cables as described in 8-56-4 "Removal Procedure" on
page 8-172
8-55-4 Removal procedure
1.) Draw the DRFG out of the Nest Frame.
2.) Remove the nut (GPN:5549432) and washer (GPN:5549433).
Figure 8-241 Drawing the DRFG
Figure 8-242 Removing the nut and washer