GE DRAFT VOLUSON™ P8/VOLUSON™P6
DIRECTION 5775469, R
EVISION 3 DRAFT (JULY 19, 2018) BASIC SERVICE MANUAL
Chapter 8 - Replacement Procedures 8-109
5.) Unscrew 2 screws on trackball and disconnect the cable side of the trackball. Refer to the figures
below.
8-33-4 Installation Procedure
1.) Put the trackball guide ring. Refer to the figure below.
- If the trackball guide ring is installed already, please skip step 1).
Figure 8-161 Unscrewing 2 screws on trackball
Figure 8-162 Installing the trackball guide ring