GE RAFT VOLUSON™ P8/VOLUSON™P6
DIRECTION 5775469, R
EVISION 3 DRAFT (JULY 19, 2018) BASIC SERVICE MANUAL
8-62 Section 8-20 - Replacement of OPIO Plastic upper cover assy
Section 8-20
Replacement of OPIO Plastic upper cover assy
8-20-1 Manpower
One person, 30 minutes
8-20-2 Tools
Philips screwdriver
8-20-3 Disassembly procedure
1.) Unscrew 9 screws at the bottom of the keyboard body. Refer to the figures below.
Figure 8-82 Unscrewing 9 screws