EVISION 3 DRAFT (JULY 19, 2018) BASIC SERVICE MANUAL
8-22 Section 8-4 - Image Settings Only - Loading Procedure
Section 8-4
Image Settings Only - Loading Procedure
8-4-1 Introduction
The Image Settings contain:
• Application Settings
• 2D Factory and 2D user Presets
• 3D/4D Factory and 3D/4D User Presets
• Annotation Presets
• Scan Assistant Configuration
• Measure configuration
• Biopsy Lines
8-4-2 Loading Procedure
see: 4-5-2 "Load Small Backup (Scan Settings)" on page 4-14
Section 8-5
Full Backup (Full System Configuration) - Loading Procedure
8-5-1 Introduction
The Full Backup contains following data:
• User Settings (databases and files containing User Programs, 2D/3D/4D Programs, Auto Text
entries, gray curves and complete System settings such as language, time/date format, etc.)
• Measure Configuration (user specific measure setup settings)
• Patient Archive (database containing patient demographic data and measurements) - no images
• Options (Permanent Key that is specific for enabled software options and Demo Key)
• Image Transfer Configuration (DICOM settings e.g., DICOM servers, AE Title, Station Name, etc.)
• Network Configuration (network settings including the computer name)
• Service Platform (state of the Service Software)
8-5-2 Loading Procedure
see: 4-5-4 "Load Full System Configuration (Full Backup)" on page 4-19
Section 8-6
Image Archive - Loading Procedure
8-6-1 Introduction
A backup of the Image Archive contains the Patient Archive (database containing patient demographic
data and measurements) + images.
8-6-2 Loading Procedure
see: 4-5-6-2 "Load Image Archive" on page 4-24