EasyManuals Logo
Home>GE>Medical Equipment>Voluson P8

GE Voluson P8 Service Manual

GE Voluson P8
620 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #483 background imageLoading...
Page #483 background image
GE DRAFT VOLUSON™ P8/VOLUSON™P6
DIRECTION 5775469, R
EVISION 3 DRAFT (JULY 19, 2018) BASIC SERVICE MANUAL
Chapter 8 - Replacement Procedures 8-121
Section 8-37
Replacement of the CASTER
8-37-1 Manpower
One person, 10 minutes
8-37-2 Tools
Monkey Wrench
8-37-3 Preparations
1.) Power Off/Shutdown the system as described in 4-3-2 on page 4-4.
NOTE: Lock casters before removing them.
8-37-4 Removal Procedure
1.) Follow the removal procedure of Section 8-24 "Replacement of ASSY Bumper Base Rubber" .
2.) Unscrew the NUT. And pull out the CASTER. Refer to the figure below.
8-37-5 Installation Procedure
1.) Install the new parts in the reverse order of removal.
Figure 8-177 Unscrewing the NUT
NOTICE
Press the caster wheel locks and ensure that they keep the system from moving.

Table of Contents

Other manuals for GE Voluson P8

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE Voluson P8 and is the answer not in the manual?

GE Voluson P8 Specifications

General IconGeneral
BrandGE
ModelVoluson P8
CategoryMedical Equipment
LanguageEnglish

Related product manuals