8-144 Section 8-48 - Replacement of the DBP
Section 8-48
Replacement of the DBP
8-48-1 Manpower
One person, 30 minutes
8-48-2 Tools
Philips screwdriver, Allen Wrench
8-48-3 Preparations
1.) Power Off/Shutdown the system as described in 4-3-2 on page 4-4.
2.) Pull out the NEST as described in Section 8-28 "Replacement of the NEST" on page 8-93.
8-48-4 Removal Procedure
1.) Remove the CSI. Refer to Section 8-47 "Replacement of the CSI" on page 8-143
2.) Remove fan connector and unscrew 8 screws.
Figure 8-205 Removing fan connector and unscrewing 8 screws