GE RAFT VOLUSON™ P8/VOLUSON™P6
DIRECTION 5775469, R
EVISION 3 DRAFT (JULY 19, 2018) BASIC SERVICE MANUAL
8-170 Section 8-55 - Replacement of UVC
3.) Unscrew 3 screws and replace UVC Piggy board.
8-55-5 Installation procedure
1.) Install the new parts in the reverse order of removal procedure.
2.) Perform
a.) Connect output (Composite or S-video) on the UVC to external monitor.
b.) Turn the system on and wait for about 30 seconds. (It takes about 30 seconds for the UVC
booting).
c.) Check that image on external monitor is displayed correctly. If you connect S-video/Composite
port on the UVC to external monitor, Cropped image (only Scan area out of the main display)
will be displayed on external monitor.
Figure 8-243 Unscrewing 3 screws