GE RAFT VOLUSON™ P8/VOLUSON™P6
DIRECTION 5775469, R
EVISION 3 DRAFT (JULY 19, 2018) BASIC SERVICE MANUAL
4-4 Section 4-3 - General Procedure
4-3-2 Power Off / Shutdown
4-3-2-1 System Shutdown
1.) If not already in read mode, freeze the image.
2.) Press the ON/OFF button (see: Figure 4-2) on the control panel. Following dialog appears.
3.) Select the SHUTDOWN
button. The system performs an automatic full shutdown sequence.
NOTE: Full shut down is also performed when pressing the ON/OFF button on the control panel twice.
4.) Switch OFF the Circuit Breaker at the rear of the system.
NOTE: The mains outlet of the system for peripheral auxiliary equipment are commonly switched with the
ON/OFF button. So the auxiliary equipment need not to be switched ON/OFF separately.
!! NOTICE:
After turning off a system, wait at least 10 seconds before turning it on again.
The system may not be able to boot if power is recycled too quickly.
Figure 4-3 Shutdown, Restart, Cancel or Sleep
!! WARNING:
Disconnection of the Main Power Cable is necessary!
For Example: When repairing the system.