GE DRAFT VOLUSON™ P8/VOLUSON™P6
DIRECTION 5775469, R
EVISION 3 DRAFT (JULY 19, 2018) BASIC SERVICE MANUAL
Chapter 7 - Diagnostics/Troubleshooting 7-67
7-10-7 Network Troubleshooting
7-10-7-1 No Connection to the Network at All
1.) Check that the network cable between the VVoluson™ P8/Voluson™P6 system and the wall
network is connected and well seated in both ends. (Use a network cable that is known to be OK.)
2.) Connect a network cable between the system and a PC by either using a hub or a cross-over cable.
Try to ping from system to IP address on PC. If OK, hardware connection inside the system is OK.
7-10-7-2 GE remote service connection
The customer gets visual information about GE remote service status (shown in the status area on the
right bottom of the screen).
Gray = Cable disconnected or no network signal on a connected cable.
Green = Cable connected to a network. Does not imply
proper network settings.
Gray = Idle State Remote access is inactive.
Yellow = Access
pending
Remote Access connection is active, but Disruptive Mode and Virtual Console
Observation (VCO) is not yet enabled.
Red = Disrupted State
Remote Access is active. All processes [UL_VNC and UL_CSD] are active. In this
state the Voluson P-Series system should NOT be used clinically.
Gray + envelope
New Software Update Package is available for download and installation. Operation
see Section 8-2-4 "Software Update Package - Download/Installation Procedure" on
page 8-13.