GE RAFT VOLUSON™ P8/VOLUSON™P6
DIRECTION 5775469, R
EVISION 3 DRAFT (JULY 19, 2018) BASIC SERVICE MANUAL
10-4 Section 10-3 - Maintenance Task Schedule
Endocavity Probe
Leakage
Current Checks
Also after corrective
maintenance or as required by
your facilities QA program.
Measurement Accuracy
Checks
Also after corrective
maintenance or as required by
your facilities QA program.
Probe/
Phantom(optional)
Checks
Check axial and lateral resolution (see Basic
User Manual Technical specifications).
Check Gain and TGC changes, vary the
focus and check reaction on screen.
Also after corrective
maintenance or as required by
your facilities QA program.
Functional Checks of all
probes Section 10-6-2
on page 10-9
Also after corrective
maintenance or as required by
your facilities QA program.
Table 10-2 Customer Care Schedule
Item Service at Indicated Time D W M A Notes

Table of Contents

Other manuals for GE Voluson P8

Question and Answer IconNeed help?

Do you have a question about the GE Voluson P8 and is the answer not in the manual?

GE Voluson P8 Specifications

General IconGeneral
CategoryUltrasound System
Probe Ports3 active probe ports
DICOMYes
Power Requirements100-240 VAC, 50/60 Hz
ApplicationsObstetrics, Gynecology
Imaging ModesColor Doppler, Power Doppler
ConnectivityEthernet, USB
Operating SystemWindows-based

Related product manuals