GE DRAFT VOLUSON™ P8/VOLUSON™P6
DIRECTION 5775469, R
EVISION 3 DRAFT (JULY 19, 2018) BASIC SERVICE MANUAL
Chapter 3 - Setup Instructions 3-29
NOTE: The mains outlet of the system for peripheral auxiliary equipment are commonly switched with the
ON/OFF button. So the auxiliary equipment need not to be switched ON/OFF separately.
5.) After complete power down, disconnect the main power cable from the system or unplug it from the
AC wall outlet socket.
6.) Press on the brakes to block the front caster wheels.
7.) Disconnect probes. (Turn the probe locking handle counterclockwise and then pull the connector
straight out of the probe port.)
WARNINGWARNING
!! WARNING:
Disconnection of the Main Power Cable is necessary!
For Example: When repairing the system.
Figure 3-18 Circuit and main power at rear of system
CAUTION
!! CAUTION:
DO NOT disconnect a probe while running (Live Scan “Write” mode)!
A software error may occur. In this case switch the unit OFF (perform a reset).
Circuit
Main Power

Table of Contents

Other manuals for GE Voluson P8

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE Voluson P8 and is the answer not in the manual?

GE Voluson P8 Specifications

General IconGeneral
CategoryUltrasound System
Probe Ports3 active probe ports
DICOMYes
Power Requirements100-240 VAC, 50/60 Hz
ApplicationsObstetrics, Gynecology
Imaging ModesColor Doppler, Power Doppler
ConnectivityEthernet, USB
Operating SystemWindows-based

Related product manuals