Model 6517B Electrometer Reference Manual Section 12: Common commands
6517B-901-01 Rev. C / August 2015 12-3
Common command descriptions
*CLS (clear status)
This command clears status registers and the error queue.
Details
The *CLS command is used to clear (reset to 0) the bits of the following registers in the Model 6517B:
ï‚· Standard event status register
ï‚· Operation event register
ï‚· Error queue trigger event register
ï‚· Sequence event register
ï‚· Arm event status register
ï‚· Measurement event register
ï‚· Questionable event register
This command also forces the instrument into the operation complete command idle state and
operation complete query idle state.
Also see
None