Model 6517B Electrometer Reference Manual Section 14: SCPI command reference
6517B-901-01 Rev. C / August 2015 14-9
[:DATA]?
Query lower limit

:SOURce <NRf>
Specify digital output (0 to 15) 0

:SOURce?
Query source (digital output value)

:STATe <b>
Enable or disable limit test OFF

:STATe?
Query state of limit test

:FAIL?
Query test result (1 = pass, 0 = fail)

:CLEAR
Path to clear failed test:

[:IMMediate]
Clear failed test indication

:AUTO <b>
Enable or disable auto clear ON

:AUTO?
Query auto clear

:CALCulate3
Subsystem to control CALC 3 (limit tests):

:PASS
Path to define "pass" digital output pattern:
:SOURce <NRf>
Specify digital output (0 to 15) 0
:SOURce?
Query digital output value
:CLIMits
Command path for composite limits:

:FAIL?
Query composite result of LIMIT 1 and

LIMIT 2 tests (1 = pass, 0 = fail)
:BSTRobe
Path to control limit-binning strobe:
:STATe <b>
Enable or disable strobe OFF
:STATe?
Query state of binning strobe
:IMMediate
Re-perform limit tests

Note: ON is the *RST default parameter, and OFF is the :SYSTem:PRESet default.
DISPlay command summary
Command Description
Default
parameter
SCPI
:DISPlay
[:WINDow[1]]
Path to locate message to top display

:ATTRibutes?
Query attributes of display characters;
1=blinking character, 0=other
:TEXT
Path to control user text messages (Note 1)

:DATA <a>
Define ASCII message "a" (up to 20
characters)

:DATA?
Query text message

:STATe <b>
Enable or disable message mode (Note 2)

:STATe?
Query text message state

:DATA?
Read data on top portion of display
:WINDow2
Path to locate message to bottom display

:ATTRibutes?
Query attributes of display characters;
1=blinking character, 0=other