Section 14: SCPI command reference Model 6517B Electrometer Reference Manual
14-92 6517B-901-01 Rev. C / August 2015
:LIMit:STATe <b>
This command is used to control the voltage limit.
Usage
:SOURce:VOLTage:LIMit:STATe <b>
:STATe?
<b>
ï‚· 0 or OFF: sable voltage limit
ï‚· 1 or ON: Enable voltage limit
Details
This command is used to enable or disable voltage limit. When enabled, the programmed voltage
limit is in effect. When disabled, voltage limit is determined by the selected range.
Attempts to enable voltage limit when the present V-source level (amplitude) exceeds the limit value
generates a Settings Conflict error.
Also see
None
:MCONnect <b>
This command is used to control the LO connection.
Usage
:SOURce:VOLTage:MCONnect <b>
:MCONnect?
<b>
ï‚· ON or 1: Enable LO connection
ï‚· OFF or 0: Disable LO connection
Details
This command is used to control the internal V-source LO to Ammeter LO connection. When enabled,
V-source LO is connected to Ammeter LO. This connection simplifies the connection requirements for
ohms measurements and source voltage measure current measurements. When disabled, this
connection is broken.
Also see
None

Table of Contents

Other manuals for Keithley 6517B

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 6517B and is the answer not in the manual?

Keithley 6517B Specifications

General IconGeneral
BrandKeithley
Model6517B
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals