Model 6517B Electrometer Reference Manual Section 14: SCPI command reference
6517B-901-01 Rev. C / August 2015 14-55
:BORDer <name>
This command specifies the binary byte order.
Usage
:FORMat:BORDer <name>
:BORDer?
<name>
ï‚· NORMal: Normal byte order for binary formats
ï‚· SWAPped: Reverse byte order for binary formats
Details
This command is used to control the byte order for the IEEE754 binary formats. For normal byte
order, the data format for each element is sent as follows:
Byte 1 Byte 2 Byte 3 Byte 4 (Single precision)
Byte 1 Byte 2 ... Byte 8 (Double precision)
For reverse byte order, the data format for each element is sent as follows:
Byte 4 Byte 3 Byte 2 Byte 1 (Single precision)
Byte 8 Byte 7 ... Byte 1 (Double precision)
The "#,0" Header is not affected by this command. The Header is always sent at the beginning of the
data string for each measurement conversion.
The ASCII data format can only be sent in the normal byte order. The SWAPped selection is ignored
when the ASCII format is selected.
Also see
None
OUTPut subsystems
The Output subsystems are used to control the V-source (standby or operate) and set polarities for the digital
output port. Commands for these subsystems are summarized in OUTPut command summary (on page 14-11).

Table of Contents

Other manuals for Keithley 6517B

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 6517B and is the answer not in the manual?

Keithley 6517B Specifications

General IconGeneral
BrandKeithley
Model6517B
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals