EasyManuals Logo
Home>Keithley>Measuring Instruments>6517B

Keithley 6517B Reference Manual

Keithley 6517B
498 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #374 background imageLoading...
Page #374 background image
Model 6517B Electrometer Reference Manual Section 14: SCPI command reference
6517B-901-01 Rev. C / August 2015 14-123
:POSetup <name>
This command programs power-on defaults.
Usage
SYSTem:POSetup <name>
SYSTem:POSetup?
<name>
ï‚· RST: Select *RST defaults on power-up
ï‚· PRESet: Select :SYSTem:PRESet defaults on power-up
ï‚· SAVX: Select saved defaults on power-up where X = 0 to 9
Details
This command is used to select or query the power-on defaults. With RST selected, the instrument
powers up to the *RST default conditions. With PRESet selected, the instrument powers up to the
:SYStem:PRESet default conditions. Default conditions are listed in the SCPI tables (refer to the
SCPI command summary (on page 14-6)).
With one of the SAV parameters selected, the instrument powers-on to the setup that is saved in the
specified memory location using the *SAV command.
Also see
:PRESet (on page 14-122)
:VERSion?
This command reads the version of the SCPI standard.
Usage
:SYSTem:VERSion?
Details
This query command is used to read the version of the SCPI standard being used by the Model
6517B.
Example
SYST:VERS
Example output:
1996.0
The above response message indicates the
version of the SCPI standard.
Also see
None

Table of Contents

Other manuals for Keithley 6517B

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 6517B and is the answer not in the manual?

Keithley 6517B Specifications

General IconGeneral
BrandKeithley
Model6517B
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals