Model 6517B Electrometer Reference Manual Section 14: SCPI command reference
6517B-901-01 Rev. C / August 2015 14-93
:CURRent commands
:RLIMit:STATe <b>
This command is used to control the resistance current.
Usage
:SOURce:CURRent:RLIMit:STATe <b>
:STATe?
<b>
0 or OFF: Disable resistance current limit
1 or ON: Enable resistance current limit
Details
This command is used to enable or disable resistance current limit for the V-source. When enabled, a
1 M resistor is placed in series with the HI V-SOURCE OUT lead to extend the current limit
capabilities of the V-source. When disabled, normal current limit occurs at 1 mA.
Also see
None
:LIMit[:STATe]?
This command is used to query the current compliance.
Usage
:SOURce:CURRent:LIMit[:STATe]?
Details
Description This command is used to determine if the V-source has reached its current compliance
limit of 1mA. The response message is explained as follows:
0 = Current compliance not reached
1 = Current compliance reached
Also see
None
:ROUTe subsystem
The commands in this subsystem are used to configure and control switching and are summarized in
ROUTe command summary (on page 14-11).

Table of Contents

Other manuals for Keithley 6517B

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 6517B and is the answer not in the manual?

Keithley 6517B Specifications

General IconGeneral
BrandKeithley
Model6517B
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals