Model 6517B Electrometer Reference Manual Section 14: SCPI command reference
6517B-901-01 Rev. C / August 2015 14-63
[:UPPer] <n>
This command is used to set the measurement range for volts, amps, and coulombs.
Usage
[:SENSe[1]]:VOLTage[:DC]:RANGe[:UPPer] <n>
[:SENSe[1]]:CURRent[:DC]:RANGe[:UPPer] <n>
[:SENSe[1]]:CHARge:RANGe[:UPPer] <n>
:RANGe[:UPPer]?
:RANGe[:UPPer]? DEFault
:RANGe[:UPPer]? MINimum
:RANGe[:UPPer]? MAXimum
<n>
0 to 210: Expected reading in volts
0 to 21e-3: Expected reading in amps
0 to 2.1e-6: Expected reading in coulombs
DEFault
21e-3 (amps)
2.1e-6 (coulombs)
MINimum: 0 (all functions)
MAXimum: Same as DEFault
Details
This command is used to manually select the measurement range for the specified measurement
function. The range is selected by specifying the expected reading as an absolute value. The Model
6517B then goes to the most sensitive range that accommodate the expected reading. For example,
if you expect a reading of approximately 10 mA, simply let the parameter (<n>) = 0.01 (or 10e-3) in
order to select the 20 mA range.
Also see
None

Table of Contents

Other manuals for Keithley 6517B

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 6517B and is the answer not in the manual?

Keithley 6517B Specifications

General IconGeneral
Input Impedance>200 TΩ
Electrometer TypeElectrometer/High Resistance Meter
Current Measurement Range20 fA to 20 mA
DisplayVacuum Fluorescent Display
Communication InterfacesIEEE-488 (GPIB), RS-232
Power Supply100 V / 120 V / 220 V / 240 V; 50 Hz / 60 Hz
Resistance Measurement RangeUp to 200 TΩ

Related product manuals