Model 6517B Electrometer Reference Manual Section 14: SCPI command reference
6517B-901-01 Rev. C / August 2015 14-137
:TRACe|:DATA
The bar (|) indicates that :TRACe or :DATA can be used as the root command for this subsystem.
From this point on, the documentation in this manual uses :TRACe. If you prefer to use :DATA,
simply replace all the :TRACe command words with :DATA.
:CLEar
This command is used to clear the buffer.
Usage
:TRACEe:CLEar
Details
This action command is used to clear the buffer of readings. If you do not clear the buffer, a
subsequent store overwrites the old readings. If the subsequent store is aborted before the buffer
becomes full, you could end up with some old readings still in the buffer.
Also see
None
:FREE?
This command is used to read the status of the memory.
Usage
:TRACe:FREE?
Details
This command is used to read the status of storage memory. After sending this command and
addressing the Model 6517B to talk, two values separated by commas are sent to the computer. The
first value indicates how many bytes of memory are available, and the second value indicates how
many bytes are reserved to store readings.
Also see
None

Table of Contents

Other manuals for Keithley 6517B

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 6517B and is the answer not in the manual?

Keithley 6517B Specifications

General IconGeneral
BrandKeithley
Model6517B
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals