Model 6517B Electrometer Reference Manual Section 14: SCPI command reference
6517B-901-01 Rev. C / August 2015 14-157
:TYPE <name>
These commands select and query a test sequence.
Usage
:TSEQuence:TYPE <name>
:TYPE?
<name>
ï‚· DLEakage: Diode leakage test
ï‚· CLEakage: Capacitor leakage test
ï‚· CIResistance: Cable insulation resistance test
ï‚· RVCoefficient: Resistor voltage coefficient test
ï‚· SRESistivity: Surface resistivity test
ï‚· VRESistivity: Volume resistivity test
ï‚· SIResistivity: Surface insulation resistance test
ï‚· SQSWeep: Square wave sweep test
ï‚· STSWeep: Staircase sweep test
ï‚· ALTPolarity: Alternating polarity resistance/resistivity test
Details
This command is used to select the desired test sequence. When test sequence is started, the
selected test runs according to how it is configured. The commands in the following paragraphs
configure the test sequences.
:TYPE? queries the selected test sequence.
Also see
:TSOurce <name> (on page 14-158)