Model 6517B Electrometer Reference Manual Section 14: SCPI command reference
6517B-901-01 Rev. C / August 2015 14-131
A/D controls
:ARSPeed <name>
This command controls autoranging speed.
Usage
:SYSTem:ARSPeed <name>
:ARSPeed?
<name>
ï‚· FAST: Fast autoranging
ï‚· NORMal: Normal autoranging
Details
This command is used to control autoranging speed. If the autoranging causes the instrument to
rapidly alternate between ranges, changing the speed to NORMal delays measurements after
changing ranges to help avoid oscillating between ranges.
Also see
None
:TSControl <b>
This command controls external temperature readings.
Usage
:SYSTem:TSControl <b>
:TSControl?
<b>
ï‚· 0 or OFF: Disable external temperature readings
ï‚· 1 or ON: Enable external temperature readings
Details
This command controls external temperature readings. When enabled (and the Model 6517-TP is
connected to the instrument), the Model 6517B makes external temperature readings. Note that
ETEMperature has to be a selected data element in order to include the temperature reading in the
data string.
Also see
ETEMperature

Table of Contents

Other manuals for Keithley 6517B

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 6517B and is the answer not in the manual?

Keithley 6517B Specifications

General IconGeneral
BrandKeithley
Model6517B
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals