EasyManuals Logo
Home>Keithley>Measuring Instruments>6517B

Keithley 6517B Reference Manual

Keithley 6517B
498 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #446 background imageLoading...
Page #446 background image
Model 6517B Electrometer Reference Manual Section 15: Calibration procedure
6517B-901-01 Rev. C / August 2015 15-23
Temperature calibration commands
This is a list of all Model 6517B temperature calibration commands.
Temperature calibration commands
Command Description
:CAL:PROT:TZERO
Calibrate the temperature input zero
:CAL:PROT:T100 0.004096
Calibrate the temperature 100 °C range
:CAL:PROT:TN25 -0.000968
Calibrate the temperature -25 °C range
Miscellaneous calibration commands
This is a list of the remaining Model 6517B calibration commands.
Miscellaneous calibration commands
Command Description
:CAL:PROT:COUNT?
Query the number of times the instrument has been
calibrated.
:CAL:PROT:DATE yyyy,mm,dd
Set the calibration date
:CAL:PROT:DATE?
Query the calibration date
:CAL:PROT:NDUE yyyy,mm,dd
Set the calibration due date
:CAL:PROT:NDUE?
Query the next calibration due date
:CAL:PROT:CODE ‘KInnnnnn’
Unlock the calibration using the password (default is
006517)
:CAL:PROT:SAVE
Save the calibration
:CAL:PROT:LOCK
Lock the calibration

Table of Contents

Other manuals for Keithley 6517B

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 6517B and is the answer not in the manual?

Keithley 6517B Specifications

General IconGeneral
BrandKeithley
Model6517B
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals