Model 6517B Electrometer Reference Manual Section 14: SCPI command reference
6517B-901-01 Rev. C / August 2015 14-81
[:STATe] <b>
This command is used to control the median volts, amps, ohms, and coulombs.
Usage
[:STATe] <b>
[:SENSe[1]]:VOLTage[:DC]:MEDian[:STATe] <b>
[:SENSe[1]]:CURRent[:DC]:MEDian[:STATe] <b>
[:SENSe[1]]:RESistance:MEDian[:STATe] <b>
[:SENSe[1]]:CHARge:MEDian[:STATe] <b>
[STATe]?
<b>
ï‚· 1 or ON: Enable median filter
ï‚· 0 or OFF: Disable median filter
Details
This command is used to control (on or off) the median filter. When enabled, the instrument determines and
displays the median reading for the specified number of readings. The :RANK command is used to specify the
number of readings used to determine the median.
Note that the median filter can be used with the average filter. When both of these filters are enabled, each
displayed reading reflects the median for the average readings.
Setting STATe to OFF with AVER:TYPE set to NONE sets AVER:STAT to OFF.
Also see
None