Section 14: SCPI command reference Model 6517B Electrometer Reference Manual
14-162 6517B-901-01 Rev. C / August 2015
:SVOLtage <NRf>
This command is used to specify the bias voltage for the specified test.
Usage
:TSEQuence:CLEakage:SVOLtage <NRf>
:TSEQuence:CIResistance:SVOLtage <NRf>
:TSEQuence:RVCoefficient:SVOLtage[1] <NRf>
:TSEQuence:RVCoefficient:SVOLtage2 <NRf>
:TSEQuence:SRESistivity:SVOLtage <NRf>
:TSEQuence:VRESistivity:SVOLtage <NRf>
:TSEQuence:SIResistance:SVOLtage <NRf>
:SVOLtage?
<NRf>
-100.0 to 100.0 Volts; 100 V range
-1000 to 1000 Volts; 1000 V range
Details
This command is used to specify the bias voltage for the specified test. Note that the resistor voltage
coefficient test uses two bias voltages; bias voltage 1 and bias voltage 2.
This configuration command is used for the following tests:
ï‚· Capacitor leakage test
ï‚· Cable insulation resistance test
ï‚· Resistor voltage coefficient test
ï‚· Surface resistivity test
ï‚· Volume resistivity test
ï‚· Surface insulation resistance test
:TSEQuence:CLEakage:SVOLtage specifies the bias voltage for the capacitor leakage test.
:TSEQuence:CIResistance:SVOLtage specifies the bias voltage for the cable insulation
resistance test.
:TSEQuence:RVCoefficient:SVOLtage[1] specifies bias voltage 1 for the resistor voltage
coefficient test.
:TSEQuence:RVCoefficient:SVOLtage2 specifies bias voltage 2 for the resistor voltage
coefficient test.
:TSEQuence:SRESistivity:SVOLtage specifies the bias voltage for the surface resistivity test.
:TSEQuence:VRESistivity:SVOLtage specifies the bias voltage for the volume resistivity test.
:TSEQuence:SIResistance:SVOLtage specifies the bias voltage for the surface insulation
resistivity test.
:SVOLtage? queries the bias voltage.
Also see