Section 14: SCPI command reference Model 6517B Electrometer Reference Manual
14-146 6517B-901-01 Rev. C / August 2015
:POFLag <name>
This command is used to control the no-operation-pending flag.
Usage
:INITiate:POFLag <name>
:POFLag?
<name>
ï‚· INCLude: Include no-operation-pending flag
ï‚· EXCLude: Exclude no-operation-pending flag
Details
This command is used to include or exclude the no-operation-pending flag in the internal process of
determining whether all pending operations of an initiate command are completed.
Also see
*OPC (operation complete) (on page 12-8)
*OPC? (operation complete query) (on page 12-9)
:ABORt
This command is used to abort the operation.
Usage
:ABORt
Details
When this action command is sent, the Model 6517B aborts the operation and returns to the top of
the trigger model. If continuous initiation is disabled, the instrument goes the idle state. If continuous
initiation is enabled, operation continues on into Arm Layer 1.
The abort command resets the source bypasses of the trigger model. If scanning, the scan pointer
resets back to the first channel in the scan list.
Also see
None

Table of Contents

Other manuals for Keithley 6517B

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 6517B and is the answer not in the manual?

Keithley 6517B Specifications

General IconGeneral
BrandKeithley
Model6517B
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals