Model 6517B Electrometer Reference Manual Section 12: Common commands
6517B-901-01 Rev. C / August 2015 12-13
*STB? (status byte query)
This command reads the status byte register.
Details
The *STB? query command is used to acquire the value (in decimal) of the status byte register. The
status byte register is shown in the following table. The binary equivalent of the decimal value
determines which bits in the register are set.
All bits, except Bit 6, in this register are set by other event registers and queues. Bit 6 sets when one
or more enabled conditions occur.
The *STB? query command does not clear the status byte register. This register can only be cleared
by clearing the related registers and queues. Register and queue structure are explained in Status
structure (on page 13-1).
For example, for an acquired decimal value of 48, the binary equivalent is 00110000. This binary
value indicates that Bits 4 and 5 of the status byte register are set.
Status bye register
Bit position B7 B6 B5 B4 B3 B2 B1 B0
Event
OSB MSS,
RQS
ESB MAV QSB EAV — MSB
Decimal
Weighting
128
(2
7
)
64
(2
6
)
32
(2
5
)
16
(2
4
)
8
(2
3
)
4
(2
2
)
—
1
(2
0
)
Value
0/1 0/1 0/1 0/1 0/1 0/1 — 0/1
Values are:
ï‚· 1 = Event Bit Set
ï‚· 0 = Event Bit Cleared
Events are:
ï‚· OSB = Operation Summary Bit
ï‚· MSS = Master Summary Status
ï‚· RQS = Request Service
ï‚· ESB = Event Summary Bit
ï‚· MAV = Message Available
ï‚· QSB = Questionable Summary Bit
ï‚· EAV = Error Available
ï‚· MSB = Measurement Summary BitWhere:
The bits of the status byte register are described as follows:
Bit 0, measurement status (MSB): Set bit indicates that a measurement event has occurred. The
event can be identified by reading the measurement event status register using the
:STATus:MEASurement? command (see SCPI command reference (on page 14-1)).
Bit 1: Not used.

Table of Contents

Other manuals for Keithley 6517B

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 6517B and is the answer not in the manual?

Keithley 6517B Specifications

General IconGeneral
BrandKeithley
Model6517B
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals