Section 15: Calibration procedure Model 6517B Electrometer Reference Manual
15-20 6517B-901-01 Rev. C / August 2015
Current calibration commands
This is a list of all current calibration commands. For all current ranges, the actual current supplied
must be within ± 5 % of the nominal value otherwise a calibration error occurs.
Current calibration commands
Command Description
:CAL:PROT:AZERO20p
Calibrate the 20 pA range zero
:CAL:PROT:A20P <+20 pA value>
Calibrate the +20 pA range
:CAL:PROT:AN20P <-20 pA value>
Calibrate the -20 pA range
:CAL:PROT:AZERO200P
Calibrate the 200 pA range zero
:CAL:PROT:A200P <+200 pA value>
Calibrate the +200 pA range
:CAL:PROT:AN200P <-200 pA value>
Calibrate the -200 pA range
:CAL:PROT:AZERO2N
Calibrate the 2 nA range zero
:CAL:PROT:A2N <+2 nA value>
Calibrate the +2 nA range
:CAL:PROT:AN2N <-2 nA value>
Calibrate the -2 nA range
:CAL:PROT:AZERO20N
Calibrate the 20 nA range zero
:CAL:PROT:A20N <+20 nA value>
Calibrate the +20 nA range
:CAL:PROT:AN20N <-20 nA value>
Calibrate the -20 nA range
:CAL:PROT:AZERO200N
Calibrate the 200 nA range zero
:CAL:PROT:A200N <+200 nA value>
Calibrate the +200 nA range
:CAL:PROT:AN200N <-200 nA value>
Calibrate the -200 nA range
:CAL:PROT:AZERO2U
Calibrate the 2 mA range zero
:CAL:PROT:A2U <+2 uA value>
Calibrate the +2 mA range
:CAL:PROT:AN2U <-2 uA value>
Calibrate the -2 mA range
:CAL:PROT:AZERO20U
Calibrate the 20 mA range zero
:CAL:PROT:A20U <+20 uA value>
Calibrate the +20 mA range
:CAL:PROT:AN20U <-20 uA value>
Calibrate the -20 mA range
:CAL:PROT:AZERO200U
Calibrate the 200 mA range zero
:CAL:PROT:A200U <+200 uA value>
Calibrate the +200 mA range
:CAL:PROT:AN200U <-200 uA value>
Calibrate the -200 mA range
:CAL:PROT:AZERO2M
Calibrate the 2 mA range zero
:CAL:PROT:A2M <+2 mA value>
Calibrate the +2 mA range
:CAL:PROT:AN2M <-2 mA value>
Calibrate the -2 mA range

Table of Contents

Other manuals for Keithley 6517B

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 6517B and is the answer not in the manual?

Keithley 6517B Specifications

General IconGeneral
BrandKeithley
Model6517B
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals