Model 6517B Electrometer Reference Manual Section 14: SCPI command reference
6517B-901-01 Rev. C / August 2015 14-143
:LAST?
This command is used to read the last alternating polarity reading stored in the buffer.
Usage
:TRACe:LAST?
Details
When this query command is sent and the Model 6517B is addressed to talk, the last reading stored
in the buffer from the alternating polarity resistance/resistivity test sequence is sent to the computer.
The format that the reading is sent over the bus is controlled from the :FORMat subsystem.
Note that this command only works while the alternating polarity test sequence is running or waiting
for a control source event to start the test sequence.
Also see
None
:TSTamp:FORMat <name>
This command is used to select the time stamp format.
Usage
:TRACe:TSTamp:FORMat <name>
:FORMat?
<name>
ï‚· ABSolute: Reference to first buffer reading
ï‚· DELTa: Time between buffer readings
Details
This command is used to select the time stamp format for buffer readings. With ABSolute selected, each time
stamp is referenced to the first reading stored in the buffer. With DELTa selected, time stamps provide the time
between each buffer reading.
Also see
:ELEMents <item list> (on page 14-53)
:TSTamp commands (on page 14-127)
:TYPE <name> (on page 14-157)