Section 14: SCPI command reference Model 6517B Electrometer Reference Manual
14-28 6517B-901-01 Rev. C / August 2015
Command Description Default parameter SCPI
:ILINe <NRf>
Select input line (1 to 6) 2
:ILINe?
Query input line
:OLINe <NRf>
Select output line (1 to 6) 1
:OLINe?
Query output line
:SSYNchronous
Path to configure semi-synchronous
trigger link:
:LINE <NRf>
Select trigger line (1 to 6) 1
:LINE?
Query trigger line
Notes:
1. Defaults for continuous initiation:
:SYSTem:PRESet enables continuous initiation.
*RST disables continuous initiation.
2. TLINK and RTCLock (Arm Layer 1) are non-SCPI parameters for the :SOURce command. The
command (:SOURce), and all the other parameters are SCPI confirmed.
3. Defaults for count (Arm Layer 2 and trigger layer):
:SYSTem:PRESet sets the count to INF (infinite).
*RST sets the count to 1.