Model 6517B Electrometer Reference Manual Section 14: SCPI command reference
6517B-901-01 Rev. C / August 2015 14-117
Measurement event Negative transition effect on measurement event register
Reading overflow Sets B0 when a reading has gone from overflow to normal
Low limit 1 Sets B1 when the reading is no longer less than the low limit 1 setting
High limit 1 Sets B2 when the reading is no longer greater than the low limit 1 setting
Low limit 2 Sets B3 when the reading is no longer less than the low limit 2 setting
High limit 2 Sets B4 when the reading is no longer greater than the high limit 2 setting
Reading available Sets B5 when the next reading is being taken
Reading underflow Sets B6 when the reading underflow condition has cleared
Buffer available Sets B7 when the trace buffer has become empty
Buffer half-full Sets B8 when the trace buffer is no longer half-full
Buffer full Sets B9 when the trace buffer is no longer full
Sequence reading
available
Sets B10 when an alternating polarity measurement is being read
Buffer pretriggered Sets B11 when the trace buffer pretrigger operations are completed
Out of limits error Sets B12 when the out of limits error has cleared
Fixture lid closed Sets B13 when fixture lid is opened
V-source compliance Sets B14 when V-source comes out of compliance
Negative transition effect on questionable event register
Questionable event Negative transition effect on questionable event register
Volts summary Sets B0 when a subsequent valid volts measurement occurs
Amps summary Sets B1 when a subsequent valid amps measurement occurs
Temperature summary Sets B4 when a subsequent valid external temperature reading occurs
Humidity summary Sets B9 when a subsequent valid relative humidity reading occurs
Ohms summary Sets B10 when a subsequent valid ohms measurement occurs
Coulombs summary Sets B11 when a subsequent valid coulombs measurement occurs
Sequence test summary Sets B12 when the sequence test is re-started
Negative transition effect on operation event register
Operation event Negative transition effect on operation event register
Calibrating Sets B0 at the end of calibration
Trigger layer Sets B5 when leaving the trigger layer
Arm layer Sets B6 when leaving an arm layer
Calculating Sets B9 when the reading calculation is completed
Idle Sets B10 when leaving the idle state
Sequence test running Sets B11 when the aborted test is restarted
Negative transition effect on trigger event register
Trigger event Negative transition effect on trigger event register
Sequence 1 Sets B1 when leaving the trigger layer
Negative transition effect on arm event register
Arm event Negative transition effect on arm event register

Table of Contents

Other manuals for Keithley 6517B

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 6517B and is the answer not in the manual?

Keithley 6517B Specifications

General IconGeneral
BrandKeithley
Model6517B
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals