Model 6517B Electrometer Reference Manual Section 15: Calibration procedure
6517B-901-01 Rev. C / August 2015 15-9
Model 6517B
range/full
scale current
Negative polarity
calibration command
Positive polarity
calibration command
±20 pA
:CAL:PROT:AN20P <value> :CAL:PROT:A20P <value>
±200 pA
:CAL:PROT:AN200P <value> :CAL:PROT:A200P <value>
±2 nA
:CAL:PROT:AN2N <value> :CAL:PROT:A2N <value>
±20 nA
:CAL:PROT:AN20N <value> :CAL:PROT:A20N <value>
±200 nA
:CAL:PROT:AN200N <value> :CAL:PROT:A200N <value>
±2 mA
:CAL:PROT:AN2U <value> :CAL:PROT:A2U <value>
±20 mA
:CAL:PROT:AN20U <value> :CAL:PROT:A20U <value>
±200 mA
:CAL:PROT:AN200U <value> :CAL:PROT:A200U <value>
±2 mA
:CAL:PROT:AN2M <value> :CAL:PROT:A2M <value>
±20 mA
:CAL:PROT:AN20M <value> :CAL:PROT:A20M <value>