Trigger setups Trigger on nth Event (B-Event Scan and Horizontal Delay on)
3. SettheHorizontalDelayModebuttontoOff.
4. Set the B Event Scan button to On.
5. Click Setup button to set the values for B Event Advance mode, Start Event, End Event, and Advance
After.
Behavior
Use this sequence to trigger the instrument on the nth occurrence of the B event after the A event.
When you enable the B-Event Scan, the instrument automatically advances the B-Event value to capture a
different portion of the signal. The B-Event value is initialized to the Start Event value and is advanced to
the End Event value by one of three Advance Methods: Sequential, Random, and Toggle.
What do you want to do next?
Learn more about B-Event scan setup. (see page 792)
Learn more about other trigger types. (see page 386)
Return to the Trigger Setup control window. (see page 385)
Trigger on nth Event (Horizontal Delay Off). (see page 453)
Trigger on nth Event (Horizontal Delay On). (see page 452)
Trigger on nth Event (B-Event Scan and Horizontal Delay On). (see page 456)
Trigger on nth Event (B-Event Scan and H orizontal Delay on)
NOTE. The B-Event Scan is not available on MSO/DPO5000 Series instruments.