EasyManuals Logo
Home>Tektronix>Test Equipment>DSA70000D Series

Tektronix DSA70000D Series User Manual

Tektronix DSA70000D Series
1002 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #649 background imageLoading...
Page #649 background image
How to ? Use a predened spectral math waveform
NOTE. The cursor readout is displayed under the multipurpose readouts or in the upper right corner
of the graticule area. The cursor units will be in dB or volts for magnitude waveforms and in degrees
or radians for
those measuring phase.
g. Select V Bars, and use the m ultipurpose knobs to align the two vertical cursors to points of interest
along the hor
izontal axis of the waveform.
h. Read the frequency difference between the cursors from the Δ: readout. Read the frequency of
each cursor relative to the zero frequency point from the cursors readout.
Take a cursor measurement
Use the following procedure to take measurements with cursors:
1. To open the Cursor Controls window, press the Cursors button on the front panel, click Cursors
on the toolbar, or select Cursor Setup from the Cursors menu.
2. Select the waveform you want to measure and a Cursor Type
(see page 73) from the control
window.
3. Position the cursors with the multipurpose k nobs.
4. Read the cursor information from the on-screen Position controls.
TIP. Select the cursor mode (Tracking or Independent) in the Cursor Setup control window.
Use a predened spectral math waveform
The following procedure is an example of setting up the instrument to perform spectral analysis o f a signal.
This example uses the probe compensation signal available on the front panel of the instrument.
1. Connect the probe compensation signal to Ch 1 through a suitable probe.
2. Press Default Setup.
3. Press Autoset.
4. From the toolbar, click Vertical, and select the Chan 1 tab.
5. Click in the Offset entry box, and then use the multipurpose knobs or keypad to set the offset to 900 mV.
6. Click in the Scale entry box, and use the multipurpose knobs or keypad to set the Channel 1 Scale
to 200 mV.
7. From the toolbar, click Math, and select the Math 1 tab.
8. Click the Spectral Mag button.
The instrument sets up a predened magnitude spectral analysis waveform. Your waveform should
be similar to the one below.
DSA/DPO70000D, MSO/DPO/DSA70000C, DPO7000C, and MSO/DPO5000 Series 627

Table of Contents

Other manuals for Tektronix DSA70000D Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix DSA70000D Series and is the answer not in the manual?

Tektronix DSA70000D Series Specifications

General IconGeneral
BrandTektronix
ModelDSA70000D Series
CategoryTest Equipment
LanguageEnglish

Related product manuals