Oscilloscope Reference Incompatible acquisition features
What do you want to do next?
Learn more
about the horizontal acquisition window.
(see page 177)
Incompatible acquisition features
The following table shows which acquisition features and modes are incompatible with other features
or modes
:
Incomp
atible with
Contro
l/Feature
Explan
ation
Refere
nce Waveforms (Ref 1– 4)
Voltag
eOffset
Offset
is an acquisition control, not a
display control
Single shot acquisition
Avera
ge
Acqui
sitions continue until the
specified number of waveforms have
been acquired and averaged
Single shot acquisition
Envelope Acquisitions continue until the
specified number of waveforms have
been
acquired for the enveloped
waveform
Mea
surements
Rol
l
Mea
surements are not available until
you stop acquisitions
Envelope, Average, or WfmDB
acquisition modes
Rol
l
Usi
ng these acquisition modes will
turn Roll mode off