Oscilloscope Reference Trigger settings (Shared - Independent)
Trigger settings (Shared - Independent)
This feature determines how the instrument applies trigger levels to the input channels:
Shared. Sets the trigger level in common across a ll input channels. Changing the trigger level of one
channel changes the trigger levels of all input channels to the same level.
Independent. Allows you to set a unique trigger level for each input channel. Changing the trigger
level on one channel does not affect the trigger level setting of the other input channels.
Trigger slope
The trigger slope determines whether the instrument nds the trigger point on the rising edge, falling edge,
or both edges of the signal.
Trigger source
The trigger source determines the source of the trigger signal. The input channels are the most commonly
used trigger sources and are available for all trigger types.
NOTE. Youhavetheoptiontoadduser-dened label for analog, reference, bus, and digital sources.
DSA/DPO70000D, MSO/DPO/DSA70000C, DPO7000C, and MSO/DPO5000 Series 929

Table of Contents

Other manuals for Tektronix DSA70000D Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix DSA70000D Series and is the answer not in the manual?

Tektronix DSA70000D Series Specifications

General IconGeneral
BrandTektronix
ModelDSA70000D Series
CategoryTest Equipment
LanguageEnglish

Related product manuals