Section 10: Using a Micromanipulator 8860 Prober Model 4200A-SCS Prober and External Instrument Control
10-22 4200A-913-01 Rev. A December 2020
5. Select the Micromanipulator 8860 Prober as the model.
6. Ensure that the Number of Pins / Positioners is correct. The number of pins defined here
determines the pins that are available to assign to a switch matrix card column.
7. Select Save.
Clarius
Use Clarius to load and run the probesites project using the new KCon configuration file, which
allows you to execute the project for this prober.
On the 4200A-SCS:
1. Open Clarius.
2. Choose Select.
3. Select Projects.
4. Search for probe.
5. Drag the probesites project to the project tree.
Figure 216: probesites project tree
6. Click Run.