Section 11: Using a manual or fake prober Model 4200A-SCS Prober and External Instrument Control
11-6 4200A-913-01 Rev. A December 2020
Figure 229: Use KCon to select a prober
5. Select the Manual Prober or the Fake Prober as the model.
6. Ensure that the Number of Pins / Positioners is correct. The number of pins defined here
determines the pins that are available to assign to a switch matrix card column.
7. Select Save.
Clarius
Use Clarius to load and run the probesites project using the new KCon configuration file, which
allows you to execute the project for this prober.
On the 4200A-SCS:
1. Open Clarius.
2. Choose Select.
3. Select Projects.
4. Search for probe.
5. Drag the probesites project to the project tree.