Model 4200A-SCS Prober and External Instrument Control Section 12: Using a Cascade Summit-12000 Prober
4200A-913-01 Rev. A December 2020 12-3
The following figure shows connections between the Cascade Summit-12000 prober to the Keithley
Instruments 4200A-SCS.
Figure 235: Connection diagram
Set up probers using the Velox prober control software
For probers using the Velox prober control software:
On the probe station computer, refer to the Velox user manual and help content for information on
how to setup and configure the GPIB interface. The following settings should be made in Velox to the
GPIB configuration for operation with the 4200A:
• Device Address: 28 - Recommend using 28, as this is the default address used by the 4200A
• Response Terminator: CR - Carriage Return
• Service Request: OFF
• TSK emulation: OFF
• SCPI string responses: ON
• Legacy SCPI status byte: ON