EasyManuals Logo

Atlas Copco ACTA 4000 User Manual

Atlas Copco ACTA 4000
136 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #98 background imageLoading...
Page #98 background image
Calibrating tools and equipment ACTA 4000 User Guide
98 (136) 2008-12 9836 4171 01
7.3 Calibrating ACTA
To calibrate ACTA, we recommend contacting your local Atlas Copco Tools representative. ACTA
calibration must be done in a certified laboratory. A certificate is issued after a calibration.
Carry out ACTA reference transducer box calibration by selecting
Conf>Calibration>ACTA. When
calibrating, follow the information windows that are displayed.
If there are less than 30 days to the next calibration date, a message window appears telling the operator
that ACTA should be calibrated before the next calibration date. If the next calibration date has passed, a
message window appears telling the operator that ACTA should be calibrated.
If a complete system calibration is carried out without transducer box, the calibration date can be changed
by selecting System calibration performed in the dialog presented in Conf>Calibration>ACTA. This
action demands a password unique to each ACTA. To receive this password, contact your Atlas Copco
representative.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Atlas Copco ACTA 4000 and is the answer not in the manual?

Atlas Copco ACTA 4000 Specifications

General IconGeneral
BrandAtlas Copco
ModelACTA 4000
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals