130
Activating a Trigger at a Set Interval
2.7 Activating a Trigger at a Set Interval
Interval triggers
This section describes how to activate a trigger at a set interval.
When you set the interval trigger to [OR] or [AND], the repeat recording setting will be set to [ON]
automatically.
Interval triggers are disabled when using the Logger Utility.
> >
1
1
2
2
1
Under [Interval trigger], select the interval trigger conditions.
If you wish to give priority to the interval trigger, use the [OR] setting. (p. 131)
If you wish to give priority to other triggers, use the [AND] setting. (p. 131)
OFF

Does not use the interval trigger.
OR Uses the interval trigger based on OR logic.
AND Uses the interval trigger based on AND logic.
2
Set the interval for the interval trigger.
DD, HH:MM:SS
The trigger will simultaneously activate when measurement starts. Subsequently, the trigger will activate at
every interval trigger interval.
HIOKI LR8450A964-03
www.GlobalTestSupply.com
Find Quality Products Online at: sales@GlobalTestSupply.com

Table of Contents

Other manuals for Hioki MEMORY HiLOGGER LR8450

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Hioki MEMORY HiLOGGER LR8450 and is the answer not in the manual?

Hioki MEMORY HiLOGGER LR8450 Specifications

General IconGeneral
BrandHioki
ModelMEMORY HiLOGGER LR8450
CategoryData Loggers
LanguageEnglish

Related product manuals