Maximum Recording Times
11.11 Maximum Recording Times
The maximum amount of time during which data can be stored in the instrument’s internal bu󰀨er
memory or on its storage media can be calculated as shown below.
If saving data in binary format, the maximum recording time can be calculated using the following
formula:
Maximum recording time = Recording capacity*
1
× recording interval (s) / data size*
2
*1: For the instrument’s internal bu󰀨er memory (256 Mwords), 512 × 1024 × 1024.
*2: Data size as described in “11.9 File Size” (p. 395).
Maximum recording times (rough estimates)
Example: Measuring 30 analog channels with 2 modules (no alarm output, no waveform
calculations)
Reduce the times shown in the following table by approx. 10% since they do not take into account
the size of waveform le headers. The fewer channels are recorded, the greater the maximum
recording time.
Recording
interval
Internal bu󰀨er memory (512 MB) Z4001 (2 GB)
100 ms 10 d 8 h 38 d 18 h
200 ms 20 d 17 h 77 d 12 h
500 ms 51 d 18 h 193 d 19 h
1 s 103 d 13 h 387 d 15 h
5 s 500 d 1162 d 21 h
10 s 500 d 3876 d 8 h
Knowledge and Information
HIOKI LR8450A964-03
www.GlobalTestSupply.com
Find Quality Products Online at: sales@GlobalTestSupply.com

Table of Contents

Other manuals for Hioki MEMORY HiLOGGER LR8450

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Hioki MEMORY HiLOGGER LR8450 and is the answer not in the manual?

Hioki MEMORY HiLOGGER LR8450 Specifications

General IconGeneral
BrandHioki
ModelMEMORY HiLOGGER LR8450
CategoryData Loggers
LanguageEnglish

Related product manuals