390
Scan Timing
U8552 and LR8532
The following diagrams provide examples of typical scan timing for the U8552 and LR8532 when
using from 16 to 30 channels. If the number of channels in use is 15 or less, the scan timing is the
same as for the U8550, U8551, LR8530, and LR8531.
Example: Data refresh interval of 20 ms, measurement on for all 30 channels, wire
break detection o󰀨
Channels 1 to 30 are scanned with a duration of approximately 0.5 ms per channel.
Data refresh interval: 20 ms
Approx. 0.5 ms
CH1 CH2 CH3 CH28 CH29 CH1 CH2CH30
Example: Data refresh interval of 20 ms; measurement on for CH1 to CH10 and
CH21 to CH30; wire break detection o󰀨
Only channels for which measurement is enabled are scanned; channels for which measurement is
disabled are not scanned.
Data refresh interval: 20 ms
Approx. 0.5 ms
CH1 CH2 CH10 CH21 CH1 CH2CH22 CH30
Example: Data refresh interval of 5 s, measurement on for all 30 channels, wire
break detection o󰀨
Channels 1 to 30 are scanned with a duration of approximately 90 ms or 105 ms per channel,
depending on the power supply frequency lter setting.
Data refresh interval: 5 s
Approx. 105 ms (power supply frequency lter: 50 Hz)
CH1 CH2 CH3 CH28 CH29 CH1 CH2CH30
Approx. 90 ms (power supply frequency lter: 60 Hz)
www.GlobalTestSupply.com
Find Quality Products Online at: sales@GlobalTestSupply.com