224
Conguring Voltage Output (VOUTPUT)
8.1 Conguring Voltage Output (VOUTPUT)
This section describes how to congure voltage output for use in powering sensors.
For more information about how to connect sensors to the voltage output terminals, see “Voltage
Output Connections” in the Quick Start Manual.
> >
1
1
1
Under [Voltage output 1] and [Voltage output 2], select the voltage.
OFF
Does not generate voltage output.
5 V Outputs 5 V DC.
12 V Outputs 12 V DC.
24 V Outputs 24 V DC. (Available only under [Voltage output 1])
Select [12 V] when providing power to the optional Z2000 Humidity Sensor.
Voltage output terminal specications
Output voltage OFF, 5 V ±10%, 12 V ±10%, 24 V ±10% (user-selectable)
Supplied current Max. 100 mA
HIOKI LR8450A964-03