Conguring numerical calculations
> >
2
3
4
1
5
6
7
1
Under [Numerical calculation], set the numerical calculation function to [ON].
OFF
, ON
2
Under [Time split calculation], select the le saving method used for the auto-save.
Disable
Performs the numerical calculation using all data acquired from the measurement start
to the measurement stop, and then saves the calculation results.
Enable Segments the data set into those with a length of the set interval beginning from the
measurement start*, performs calculations using the segmented data sets, and saves
the calculation results by interval basis.
*:
When the trigger setting is enabled, segments the data set beginning from the start
trigger point.
Timed The rst interval length will automatically be adjusted so that calculation values yielded
at regular intervals ([Split time]) base on the [Reference time] are saved. (Only the
rst interval will become shorter than the [Split time].)
If the [Enable] or [Timed] setting has been selected, calculation results will be saved for each time interval.
When the numerical calculation result type for the auto-save is set to [OFF], the [Time split calculation]
setting is disabled.
See “Auto save (real-time save)” (p. 142).
Set the format for numerical calculation results to [Text (CSV)].
(When [Time split calculation] is set to [Enable])
Under [Split time], set the time interval at which to perform the calculation.
0 days 00:01 to 30 days 23:59
For example, if the segment time is set to 10 min., the calculation will be performed, and the calculation results
saved, every 10 min.
Reference
time
Segment time
Start
measurement
Stop
measurement
Calculation results will be saved at the specied intervals
based on the reference time.
The value 0 d 0 h 0 min. cannot be specied as the
segment time. The value will be automatically changed to
0 d 0 h 1 min.
www.GlobalTestSupply.com